OXYGEN FILLING AND DECOMPRESSION VALVE ADAPTER
This Oxygen Filling & Decompression Valve Adapter is designed
for all type of Potential Gums & Induction Period bombs. The NPC
210 kit provides the operator with a high level of safety. The kit has to be connected to an oxygen supply and to earth. A sliding door completely prevents the operator from any valve access when putting the test cylinder in place. The admission and exhaust pre s s u res are displayed. The filling and decompression process is achieved through a needle valve. T h i s operation is done inside the cabinet in order to protect the opera t o r. The admission pressure is 4.000 KPa and the exhaust one can be modulated from 0 to 1.600 KPa. The system is connected to an oxygen supply with a 6 mm gyrolock and with a male RBE03 STAUBLI. REF. 40930 The oxygen filling and decompression valve adapter model NPC 210 is designed for all type of potential gums and indiuction period bombs. The kit provides the operator with a high level of safety.
See also: AUTOMATIC INDUCTION PERIOD AND POTENTIAL RESIDUE
|